Procedure for torque calibration under constant rotation investigated on a nacelle test bench

Konferenz: Sensoren und Messsysteme - 19. ITG/GMA-Fachtagung
26.06.2018 - 27.06.2018 in Nürnberg, Deutschland

Tagungsband: Sensoren und Messsysteme

Seiten: 4Sprache: EnglischTyp: PDF

Persönliche VDE-Mitglieder erhalten auf diesen Artikel 10% Rabatt

Autoren:
Weidinger, Paula; Foyer, Gisa; Kumme, Rolf (Physikalisch-Technische Bundesanstalt, Braunschweig, Germany)
Kock, Stefan; Gnauert, Jonas (RWTH Aachen, Chair for Wind Power Drives, Aachen, Germany)

Inhalt:
In order to determine the efficiency of multi-MW nacelles, the torque measurement of multi-MW nacelle test benches has to be performed with an accuracy of approximately 0.2%. To this end, the torque transducer has to be traced to national standards in the MNm range. This can be done by using a new torque calibration procedure under constant rotation performed on a nacelle test bench. The results, in the form of an indication deviation including the measurement uncertainty of less than 0.22%, are discussed in this paper.