A new miniaturized Electronic Pill (ePille(R)) for mass medical application with bi-directional communication system

Konferenz: MikroSystemTechnik - KONGRESS 2007
15.10.2007 - 17.10.2007 in Dresden, Germany

Tagungsband: MikroSystemTechnik

Seiten: 4Sprache: EnglischTyp: PDF

Persönliche VDE-Mitglieder erhalten auf diesen Artikel 10% Rabatt

Autoren:
Fawaz, Nidal; Jansen, Dirk (Institute for Applied Research, Offenburg, Germany)

Inhalt:
A new electronic capsule with bidirectional communication system is being developed for multi-task application. The capsule is designed to be a platform for medical assistant application inside the body. The designed telemetry unit is a synchronous bidirectional communication block using continuous phase DQPSK of 115 kHz low carrier frequency for inductive data transmission suited for human body energy transfer. The communication system can assist the electronic pill to trigger an actuator for drug delivery, to record temperature, or to measure pH of the body. It consists additionally to a 32bit processor, memory, external peripheries, and detection facility. The complete system is designed to fit small-size mass medical application with low power consumption, size of 7x25mm. The system is designed, simulated and emulated on FPGA. A final layout of the complete chip design is still under progress.